17
Absorbance Ratio
Measuring Samples
Thermo Scientific
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
81
To measure samples automatically (using Auto 6 or Auto 3)
1. Install the blank and samples.
2. Press
Enter
.
To measure samples manually (using Manual 6 or Single Cell Holder)
1. In the Absorbance Ratio screen, press
Run Test
.
2. Install the blank and sample.
A 6-Position Cell Holder can hold five samples.
3. Press
Measure Blank
.
If a 6-Position Cell Holder is installed, it moves to the B position to measure the blank
and returns to its previous position.
4. Press
Measure Sample
.
If a 6-Position Cell Holder is installed, press the cell position buttons to reposition the cell
holder and measure the rest of the samples manually.