22
Performance Verification
Resolution (UV-Vis only)
112
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
Thermo Scientific
If the test fails, follow these guidelines:
• Repeat the test twice to verify that the test is failing consistently.
• Contact technical support for more troubleshooting advice.
Resolution (UV-Vis only)
This test measures the ability of the spectrophotometer to resolve adjacent features in a
spectrum. The test is performed using a 0.02% (v/v) solution of toluene in hexane and
requires a hexane blank.
When running the internal standard test, remember that the wavelengths and tolerance values
are preset and cannot be changed.
To run the Resolution test
1. Highlight
Resolution
and press
Enter
.
Ensure that hexane is in the blank position and toluene in hexane is in cell 1.