21
5
Cell Correction
Every test setup screen provides access to Cell Correction.
Before running Cell Correction:
• Clean the inside and outside of all the cells to be matched.
• Fill the cells with distilled water (or other blank solution), and place them in the sample
compartment (see
“Selecting and Positioning Cuvettes”
). Place the blank
cuvette in Cell “B” of the cell changer.
Cell Correction
To run Cell Correction
1. Load the test type or stored test.
2. If Cell Correction is not visible, highlight
More parameters
and press
Enter
.
Note
Cell Correction is not active from the Main (Basic ATC) screen.
Note
Cell Correction is active only when the 6-Position Cell Holder is set to either
Auto 6 or Auto 3. The feature is not active when the cell holder is set to 1-Cell or Manual
6, nor when the Single Cell Holder is installed.