The instrument takes up to 12 hours to come to thermal equilibrium. For best results, re-run
alignment and calibration after the first 12 hours of operation.
c. Run the Nicolet Summit OA - PV Test workflow. The performance verification workflow uses
reference material to run a series of standard tests to verify your instrument operation and to
ensure the accuracy of your data.
i. Return to the Dashboard.
ii. Scroll to the Workflows pane and select Nicolet Summit OA - PV Test.
iii. Right click the workflow and select Run to start the workflow.
Thermo Scientific
Nicolet Summit OA Spectrometer User Guide
55
Maintenance