Operation
Physical Inputs & Outputs
Figure 4-11.
Current Input Tabs
To access the information on the Current Input tabs using only the remote transmitter
keypad, see
The current inputs can operate in Live mode or Manual mode. In the Manual mode,
the user may enter a value in the Manual Value (EU) textbox. In Live mode, the user
does not have this option.
Once a value in engineering units have been entered into the Manual Value (EU) field,
the EZ Cal II software will convert that value into milliamps and display that value in
the Present Value field.
1.
Choose the type of input from the selections in the dropdown list.
•
Not Used
•
Temperature (Used for Density)
•
Pressure (Used for Level)
•
Flow (Used for Density)
•
Density (Used for Level)
Based on the input type selected, the fields available for configuration and the
information contained in the dropdown lists may change. The following tables
provide the units available for selection based on the various type of input.
Table 4-1.
Temperature Input, Engineering Unit
Display
Description
ºC
Degree Celsius
ºF
Degree Fahrenheit
4
-
18
LevelPRO User Manual
Thermo Scientific