Operation
Commands
5.
Totalizer Overrun Alarm
a.
The alarm is triggered when a totalizer value goes above 1,000,000.
b.
The alarm clears when the user takes action to clear the mode/fault
Tachometer Accumulation totalizer overrun flag.
6.
Calibration Aborted Alarm
a.
The alarm is triggered when the calibration is aborted.
b.
The alarm clears when the user restarts the calibration process.
7.
Detector Over Range Alarm
a.
The alarm is triggered when the IBP channel one
data count is more then
150,000.
b.
The alarm clears when the IBP channel one data count is less than 150,000.
8.
Detector Under Range Alarm
a.
The alarm is triggered when the IBP channel one data count is less than 200
b.
The alarm clears when the IBP channel one data count is more than 200.
Commands
The system commands provide the user with the ability to directly control certain
functions of the detector.
Once the desired command or commands have been selected, click the Submit button
to execute. Click the Refresh button to reset the screen or the Cancel button to exit.
To access the information on the Commands tab using only the remote transmitter
keypad, see
Common Action
The Common Actions dropdown provides the user a way to change a large group of
parameters with one command.
Figure 4-9.
Commands Screen
4
-
14
LevelPRO User Manual
Thermo Scientific