F
lash Application Firmware
B-2
LevelPRO User Manual
Thermo Scientific
Program the Main
CPU
1. Use a three-wire cable to connect PC-to- J2A of the Main CPU PCA. The cable
connections are shown in the
2. Open EZ Cal II and set the communications parameters as shown below.
Name:
Unit ID: 1
Port: Select the associated Com port
Baud rate: 57600
Parity: None
Data Bits: 8
Stop Bits: 1
Time Out (sec): 8
3. Click the green Connect button.
4. A Communication Fail pop-up window will be displayed. Click the OK button.
Figure B-2.
Communication Fail Window