Startup & the Setup Wizard
The Setup Wizard
3
-
16
LevelPRO User Manual
Thermo Scientific
5.
The Time Remaining (sec) indicator will count down the remaining sample time.
6.
Should there be a need to stop the calibration before completion, click the Abort
CAL button.
7.
The Detector Avg Count field displays live data during the calibration.
8.
Once the calibration is complete, the Detector Avg Count, CAL Temperature and
CAL/Ref from Latest textboxes will auto-populate with information gathered
during the calibration.
9.
Click to either Accept Cal or Reject Cal.
10.
Click the Next button to advance through screens of level
point data collected
during the detector calibration.
11.
Save the file to the computer or save the file to the computer and download it to
the LevelPRO detector.
Note:
To exit the Setup Wizard without saving the input data, click the
Cancel button.
Detector 1 has now been successfully set up for a level application. Once the data has
been saved, click the Home button to return to the first screen of the Setup Wizard.
Once Detector 1 has been configured for a level application and saved, there are
additional parameters that can be programmed from the EZ Cal II software.
See the section regarding
for further information.