Operation
Mode/Fault Alarm Setup
4.
System Fault Alarm
a.
The alarm is triggered when any of the following actions occur.
i.
Diagnostic error:
ii.
An Input Scan Error occurs and the detector has a problem reading any of
the following devices:
Analog inputs
Main Board Temperature
Detector Analog input
Detector RTD input
b.
The alarm clears when the error condition is cleared.
5.
Current Output A at Minimum Alarm
a.
The alarm is triggered when the value of Current Output A is below the
minimum set value.
b.
The alarm is cleared when the value of Current Output A is above the
minimum set value.
6.
Current Output A at Maximum Alarm
a.
The alarm is triggered when the value of Current Output A is above the
maximum set value.
b.
The alarm is cleared when the value of Current Output A is below the
maximum set value.
7.
Current Output B at Minimum Alarm
a.
The alarm is triggered when the value of Current Output B is below the
minimum set value.
b.
The alarm is cleared when the value of Current Output B is above the
minimum set value.
8.
Current Output B at Maximum Alarm
a.
The alarm is triggered when the value of Current Output B is above the
maximum set value.
b.
The alarm is cleared when the value of Current Output B is below the
maximum set value.
9.
Current Output C at Minimum Alarm
a.
The alarm is triggered when the value of Current Output C is below the
minimum set value.
b.
The alarm is cleared when the value of Current Output C is above the
minimum set value.
4
-
10
LevelPRO User Manual
Thermo Scientific