Operation
Application Screens
Figure 4-31.
Application Screen, Measurement Data Tab
Process Alarm
The process alarms provide a signal to the external device when the process value goes
above or below a set point value. The detector provides sixteen process alarms. The
system can be configured to setup a process alarm based on one of the four
measurements. An alarm is activated when the measurement value reaches the specified
set point. The relative values assigned to the set point and clear point determines
whether the alarm is a high alarm or low alarm. The alarm condition must exist for at
least the duration of action delay time for the alarm to trigger and take the selected
action.
To access the information on the Application Process Alarm tabs using only the remote
transmitter
4
-
50
LevelPRO User Manual