Connecting the Inlet Plumbing
Setting Up the Inlet for High-Flow Infusion
48
LTQ Series Getting Connected Guide
Thermo Scientific
Figure 30.
Connecting the union Tee to the divert/inject valve
Connecting the LC Pump to the Divert/Inject Valve
Y
To connect the LC pump to the divert/inject valve
1. Using a fingertight fitting and a ferrule, connect a length of PEEK tubing to port 2 of the
shows the ports of the divert/inject valve.
2. Using an appropriate fitting and ferrule, connect the other end of the tubing to the outlet
of the LC.
Connecting the Divert/Inject Valve to a Waste Container
Y
To connect the divert/ inject valve to a waste container
1. Using a fingertight fitting and a ferrule, connect a length of red PEEK tubing to port 1 of
the divert/inject valve.
shows the ports of the divert/inject valve.
2. Insert the other end of the tubing into a suitable waste container.
6
5
4
3
2
1
Stainless steel ferrule
(P/N 2522-3830)
Stainless steel nut
(P/N 2522-0066)
PEEK tubing
(P/N 00301-22912)
LC TEE union
Infusion line
LC union
Syringe