Maintenance
Packing for service
Thermo Fisher Scientific
Thermo Scientific KingFisher
®
Flex User Manual
71
2.
Empty the turntable.
3.
Switch OFF the power and disconnect the mains supply cable
(Figure 3–14).
4.
Disinfect the outside of the instrument using a cloth dampened
with 70% ethanol.
5.
Place the instrument in a large plastic bag. Ensure that the see-
through lid has been removed.
6.
Place a cloth soaked in the prepared solution into the bag. Ensure
that the cloth does not make contact with the instrument.
7.
Close the bag firmly and leave the instrument in the bag for at least
24 hours.
8.
Remove the instrument from the bag.
9.
Clean the instrument using a mild detergent.
10.
Remove any stains using 70% ethanol.
11.
After performing this decontamination procedure, enclose a signed
and dated Certificate of Decontamination both inside the transport
package and attached to the outside of the package (see Appendix A:
“Certificate of Decontamination”).
To pack for service, follow the guidelines presented below.
Caution
It is important that the instrument is thoroughly
decontaminated before it is removed from the laboratory or any
servicing is performed on it.
▲
When you ship the instrument for service, remember to:
●
Inform about the use of hazardous materials.
●
Decontaminate the instrument beforehand.
●
Install the transport locks.
●
Place the KingFisher Flex head into its transportation box.
●
Pack the instrument according to the enclosed packing instructions.
Packing for
service
Summary of Contents for KingFisher Flex
Page 1: ...KingFisher Flex ServiceManual Version1 1 ...
Page 2: ......
Page 4: ...1 2 Introduction Version 1 1 Service Manual Thermo Fisher Scientific KingFisher Flex ...
Page 7: ...Thermo Scientific KingFisher Flex User Manual DRAFT Rev 1 0 Cat no N07669 ...
Page 104: ...Glossary 98 Thermo Scientific KingFisher Flex User Manual Thermo Fisher Scientific ...
Page 108: ...102 Thermo Scientific KingFisher Flex User Manual Thermo Fisher Scientific ...
Page 109: ...Thermo Fisher Scientific Thermo Scientific KingFisher Flex User Manual 103 Notes ...
Page 110: ...104 Thermo Scientific KingFisher Flex User Manual Thermo Fisher Scientific Notes ...
Page 118: ...4 2 Block Diagrams Version 1 1 Service Manual Thermo Fisher Scientific KingFisher Flex ...
Page 120: ...4 4 Block Diagrams Version 1 1 Service Manual Thermo Fisher Scientific KingFisher Flex ...
Page 122: ...5 2 Component Layouts Version 1 1 Service Manual Thermo Fisher Scientific KingFisher Flex ...
Page 130: ...5 10 Component Layouts Version 1 1 Service Manual Thermo Fisher Scientific KingFisher Flex ...
Page 132: ...6 2 Schematic Diagrams Version 1 1 Service Manual Thermo Fisher Scientific KingFisher Flex ...
Page 154: ...6 24 Schematic Diagrams Version 1 1 Service Manual Thermo Fisher Scientific KingFisher Flex ...
Page 156: ...7 2 Software Version 1 1 Service Manual Thermo Fisher Scientific KingFisher Flex ...
Page 176: ...7 22 Software Version 1 1 Service Manual Thermo Fisher Scientific KingFisher Flex ...
Page 178: ...8 2 Testing and Adjusting Version 1 1 Service Manual Thermo Fisher Scientific KingFisher Flex ...
Page 198: ...10 2 Power Supply Version 1 1 Service Manual Thermo Fisher Scientific KingFisher Flex ...
Page 204: ...11 2 Wiring Diagrams Version 1 1 Service Manual Thermo Fisher Scientific KingFisher Flex ...
Page 216: ...11 14 Wiring Diagrams Version 1 1 Service Manual Thermo Fisher Scientific KingFisher Flex ...
Page 228: ...13 2 Bulletins Version 1 1 Service Manual Thermo Fisher Scientific KingFisher Flex ...