DensityPRO Specific Parameters
Parameter
Type
R/W
Description
ALARM3
ALM
R/W
Set up alarm triggering point, clear point, and gauge actions when alarm happens for
ALARM3.
Sub-parameter
Description
Select the action used to indicate that the alarm has been
triggered.
ACTION
0 = Nothing
1 = Control relay 1
2 = Control relay 2
3 = Output 1 to FAULT LOW
5 = Output 1 to FAULT HIGH
SET_POINT
The process level at which the alarm will activate
Select a clear point (clr point) or dead band (deadband) to clear
the alarm.
CLR_MODE
0 = clr point
1 = deadband
CLR_POINT_
DEADBAND
Enter clear point or dead band value.
ALARM4
ALM
R/W
Set up alarm triggering point, clear point, and gauge actions when alarm happens for
ALARM4.
Sub-parameter
Description
Select the action used to indicate that the alarm has been
triggered.
ACTION
0 = Nothing
1 = Control relay 1
2 = Control relay 2
3 = Output 1 to FAULT LOW
5 = Output 1 to FAULT HIGH
SET_POINT
The process level at which the alarm will activate
Select a clear point (clr point) or dead band (deadband) to clear
the alarm.
CLR_MODE
0 = clr point
1 = deadband
CLR_POINT_
DEADBAND
Enter clear point or dead band value.
Thermo Fisher Scientific
DensityPRO with
F
OUNDATION
™ fieldbus
Application Guide
C-5
Summary of Contents for DensityPRO
Page 2: ......
Page 3: ...DensityPRO With FOUNDATIONTM Fieldbus Application Guide P N 717917 Revision A...
Page 4: ......
Page 6: ...This page intentionally left blank...
Page 8: ...This page intentionally left blank...
Page 44: ...This page intentionally left blank...
Page 48: ...This page intentionally left blank...
Page 52: ...This page intentionally left blank...
Page 56: ...This page intentionally left blank...
Page 62: ...This page intentionally left blank...
Page 78: ...This page intentionally left blank...
Page 94: ...This page intentionally left blank...