DensityPRO Setup
Density Setup
The Density Setup menu takes you through the steps required to configure
the density measurement. These steps are listed below.
Density Setup
1.
Go to
AI_TRANSDUCER > Root Menu > Setup > 1 Density Setup.
2.
When prompted, remove the loop from automatic control.
3.
Source Head Geometry
: Scroll through list of source head models, and
select the model used on your gauge. The source head model is used to
set a “geometry factor” to tune the gauge’s performance.
4.
Material Type:
Select the correct material type: slurry, solution, single
phase, or emulsion.
5.
Carrier Gravity and Solids Gravity:
Additional menu items are
displayed to enter the required specific gravity values for that material
type as listed below.
a.
If you selected slurry, enter the carrier gravity and solids gravity.
b.
If you selected solution, enter the solvent gravity and solution type.
c.
If it is either unnecessary or impossible to describe your process
material as a slurry, emulsion, or solution, select single phase.
d.
If you selected emulsion, enter the specific gravity of the carrier
liquid (Fluid_1_gravity) and suspended liquid (Fluid_2_gravity) in
g/cc.
6.
Primary Measurement:
The available measurements depend on
selection of material type selected. The example below shows the
measurements available for slurry.
4-2
DensityPRO with
F
OUNDATION
™ fieldbus
Application Guide
Thermo Fisher Scientific
Summary of Contents for DensityPRO
Page 2: ......
Page 3: ...DensityPRO With FOUNDATIONTM Fieldbus Application Guide P N 717917 Revision A...
Page 4: ......
Page 6: ...This page intentionally left blank...
Page 8: ...This page intentionally left blank...
Page 44: ...This page intentionally left blank...
Page 48: ...This page intentionally left blank...
Page 52: ...This page intentionally left blank...
Page 56: ...This page intentionally left blank...
Page 62: ...This page intentionally left blank...
Page 78: ...This page intentionally left blank...
Page 94: ...This page intentionally left blank...