2
Installation
Installing the Standard Analytical Flow Cell
Thermo Scientific
Accela UV/Vis Detector Hardware Manual
21
b. Using the 1/8 in. ball-end hex screwdriver (supplied), secure the cover by tightening
the flow cell cover retaining screw (see
9. Close the front doors of the detector.
10. Before you start the pump flow, ensure that the outlet tubing drains to the waste
container.
Installing the Standard Analytical Flow Cell
You can order a standard 1 cm analytical flow cell from Thermo Fisher Scientific. The wetted
volume of the standard analytical flow cell is 15 μL.
To install the standard analytical flow cell
1. Turn the detector off and unplug the power cord from line power.
2. Do one of the following:
a. If the LightPipe flow cell is not installed, go to
b. If the LightPipe flow cell is installed, remove it as described in
and store it as described in
. Then go to
3. Open the flow cell compartment and remove the LightPipe flow cell retaining block as
described in
through
in
“Mounting the LightPipe Flow Cell”
.
4. Remove the LightPipe flow cell mounting assembly as follows:
a. Unscrew the tubing clamp thumbscrew (see
) and set the tubing clamp
aside.
CAUTION
To prevent possible exposure to UV light, turn the detector off and unplug
the power cord from the power receptacle on the back panel.