Troubleshooting
Detector-Related Problems
Thermo Scientific
Accela UV/Vis Detector Hardware Manual
59
Random noisy baseline Contaminated flow cell
Flush the flow cell with cleaning solvents. See
“Cleaning the LightPipe Flow Cell”
.
Check for leaks.
Leak in sample inlet line
Check all of the fittings for leaks, from the
column outlet to the flow cell inlet.
Bubble trapped in flow cell
Increase the flow rate until the bubble is
removed. Connect a backpressure device to the
flow cell (check pressure rating to avoid
rupturing the flow cell).
Leaking flow cell
Replace the flow cell.
Insufficient lamp warm-up
Allow a 30 minute warm-up for normal
operation and a 1.5 hour warm-up for maximum
sensitivity.
Aging or defective lamp
Replace the lamp.
Ground loop problem between
integrator and detector
Check the cable connections to the detector
output; do not ground at both ends of the cable.
Dirty flow cell or lamps
Clean the dirty component. See
Integrator input and output voltage not
matching
Verify that the integrator is connected to the
appropriate analog output connections on the
back panel of the detector. See
.
Check the attenuation setting on the integrator.
Excessive baseline drift Contaminated flow cell
Flush the flow cell with cleaning solvents. See
“Cleaning the LightPipe Flow Cell”
.
Check for leaks.
Mobile phase contamination
Replace the mobile phase with fresh mobile
phase that is made with high-purity solvents.
Material bleeding from column
Clean or replace the column.
Leaks in system or flow cell
Check all of the fittings for leaks. Replace the
flow cell.
Tiny bubble trapped in the flow cell
Increase the flow rate until the bubble is
removed. Connect a backpressure device to the
flow cell outlet (check backpressure rating to
avoid rupturing flow cell).
Excessive temperature fluctuations
Remove the system from drafts. Thermostatically
control the column temperature.
Table 6.
Detector-related problems (Sheet 2 of 3)
Symptom
Cause
Remedy