![Thermo Scientific Accela UV/Vis Plus Hardware Manual Download Page 59](http://html1.mh-extra.com/html/thermo-scientific/accela-uv-vis-plus/accela-uv-vis-plus_hardware-manual_1105370059.webp)
3
Routine Maintenance
LightPipe Flow Cell Maintenance
Thermo Scientific
Accela UV/Vis Detector Hardware Manual
43
3. Remove the flow cell cover.
4. Using the 3/32 in. ball-end hex screwdriver (supplied), loosen the two photodiode
assembly mounting screws three turns (
).
Figure 44.
Photodiode assembly mounting screws
5. Move the photodiode assembly forward to maximize the space behind it to install the
LightPipe flow cell.
6. Loosen and remove the thumbscrew holding the retaining bar to the flow cell retaining
block (see
CAUTION
Do not touch the surface of the photodiode. When removing the
photodiode mount, be sure to place the mount where the photodiode will not get
dirty or scratched.
Photodiode assembly