2
Installation
Making the Back Panel Connections
Thermo Scientific
Accela UV/Vis Detector Hardware Manual
27
Setting the Unit ID
The UV/Vis detector arrives with the unit ID preset using the two rotary switches located on
the back panel.
shows the rotary switches preset to a value of 01. The range of
values for the unit ID is 01 to 99. The value of 00 is reserved for special service functions.
The unit ID must correspond with the stack ID specified in the Instrument Configuration
application. Do not change the unit ID setting for your detector unless you are controlling
more than one UV/Vis detector from one computer. For information about configuring your
detector, refer to the data system Help.
Use a small flathead screwdriver to change the setting of the rotary switches.
Figure 34.
Unit ID rotary switches set to a value of 01
Connecting the System Interconnect Cable or Cables
The timing of the Accela LC modules is synchronized through a system interconnect cable or
cables. The contact-closure cabling system depends on the autosampler model.
To make the contact closure connections for your system, follow the appropriate procedure:
•
Making Contact Closure with an Accela Autosampler
•
Making Contact Closure with an Accela Open Autosampler
Making Contact Closure with an Accela Autosampler
For an LC system with an Accela Autosampler, an Accela pump (all models), an Accela
UV/Vis Detector, and an optional MS detector, use the system interconnect cable shown in
to make the contact closure connections.
1
2 3 4
5
6
7
8
9
0
1
2 3 4
5
6
7
8
9
0