2
Installation
Unpacking and Inspecting the UV/Vis Detector
Thermo Scientific
Accela UV/Vis Detector Hardware Manual
11
Unpacking and Inspecting the UV/Vis Detector
Carefully remove the UV/Vis detector from the shipping container and inspect both the
detector and the packaging for any signs of damage. If you find any damage, save the shipping
materials and immediately contact the shipping company.
In addition to the UV/Vis detector, the shipping container arrives with the following items:
• A CD or DVD that contains the documentation as PDF files (see
) and the instrument control software
• An accessory kit containing cables, tubing, fittings, and other hardware (see
)
• A LightPipe flow cell kit (see
Placement and Environmental Requirements
Place the UV/Vis detector on a benchtop as close as possible to the chromatographic
column outlet. This minimizes the length of tubing required to connect the LC column outlet
to the LightPipe flow cell inlet. If you are using other Accela modules, place the UV/Vis
detector on top of the autosampler.
Ensure that the location meets the following requirements:
• A draft-free location away from an open window, air conditioner vents, or other
circulating air sources.
• A stable room temperature (± 0.5 °C) for applications requiring maximum detection
sensitivity.
• Clearance of at least 15 cm (6 in.) between the back panel of the UV/Vis detector and any
wall or obstruction. This clear space provides access to the back-panel connectors and a
free flow of cooling air.
Note
Before you install the software driver for the UV/Vis detector, you must first install
the Thermo Scientific data system.