2
Installation
Installing the LightPipe Flow Cell
16
Accela UV/Vis Detector Hardware Manual
Thermo Scientific
4. Using a one-piece PEEK fitting, connect the (blue) 0.010 in. ID PEEK tubing (supplied)
to the flow cell outlet port. Tighten the fitting with a 1/4 in. open-end wrench.
• For the 10 mm LightPipe flow cell, see
and
• For the 50 mm LightPipe flow cell, see
and
5. Connect the other end of the (blue) 0.010 ID PEEK outlet tubing to the solvent waste
container for the LC system.
Mounting the LightPipe Flow Cell
To install the LightPipe flow cell
1. Open the front doors of the detector.
2. Remove the flow cell cover as follows:
a. Using the 1/8 in. ball-end hex screwdriver (supplied), unscrew the flow cell cover
retaining screw (see
).
Figure 16.
Flow cell cover retaining screw
b. Pull the flow cell cover forward and away from the front panel.
Note
Before you mount the LightPipe flow cell, connect the inlet and outlet tubing. See
“Connecting the Inlet and Outlet Tubing”
.
Flow cell cover
retaining screw