The Measurement Settings menu contains a number of submenus where
instrument parameters and settings can be read and modified.
The following screens show the measurement settings in single range mode
and dual or auto range mode. In the dual and auto range modes, both the
“High Range” or “Low Range ” buttons averaging buttons will be available.
Home Screen>Settings>Measurement Settings (single range mode)
Home Screen>Settings>Measurement Settings (dual or auto range
mode)
The Measurement Setting menu contains the following information:
●
Averaging Time:
Sets the averaging time for the CO
2
sample
measurement and O
2
sample measurement (if option is enabled).
●
Range Mode Selection:
User can choose between the various range
modes: single, dual, or auto. For more information, see “
Range Mode
Selection
” on page
3-77
●
Range Setting:
Sets the concentration range for the analog outputs when
in single range mode.
Measurement
Settings