Calibration
Calibration
4-8
410iQ Instruction Manual
Thermo Scientific
5.
Press the back button to back up a step to the multipoint menu.
6.
Introduce a CO
2
concentration 50% of the URL.
7.
Select
Point 2
.
8.
Enter the span concentration being introduced to the instrument. Press
Calibrate
.
The instrument will perform a series of calculations and save the new
parameters.
9.
Press the back button to back up a step to the multipoint menu.
10.
Introduce a CO
2
concentration 20% of the URL.
11.
Select
Point 3
.
12.
Enter the span concentration being introduced to the instrument. Press
Calibrate
.
The instrument will perform a series of calculations and save the new
parameters.
The instrument can also be calibrated using a dual three-point calibration.
The calibration points are divided into a “high” range and a “low” range
consisting of three points each. See “
Multipoint Calibration
” on page
4-6
.
In order to generate data of the highest confidence, it is recommended that
a multipoint calibration be performed:
●
every three months
●
any time any major disassembly of components is performed
High and Low
Multipoint
Calibration
Calibration Frequency
Summary of Contents for 410iQ
Page 1: ...410iQ Instruction Manual Carbon Dioxide Analyzer 117081 00 1Apr2020 ...
Page 2: ......
Page 18: ......
Page 35: ...Operation Instrument Display Thermo Scientific 410iQ Instruction Manual 3 9 ...
Page 160: ......
Page 171: ...Calibration Calibration Thermo Scientific 410iQ Instruction Manual 4 11 ...
Page 188: ......
Page 264: ......
Page 284: ......
Page 304: ......
Page 305: ......