Operation
Calibration
Thermo Scientific
410iQ Instruction Manual
3-25
The Multipoint Calibration screen is used to calibrate the instrument using
up to three gas concentrations (cal points) for each range. It is suggested to
use three cal-points, as this will give the most accurate readings over the
entire range.
The following screen is shown in single range mode. In dual or auto range
modes, “High Range” or “Low Range” is displayed to indicate the
calibration of the high or low coefficients. The High Range Multipoint
Calibration and Low Range Multipoint Calibration screens function the
same way.
Note
Pressing the Multipoint Calibration button responds with “It is
recommended to calibrate all three points” and requires and OK action.
▲
Home Screen>Calibration>Advanced Calibration>Multipoint
Calibration (single range mode)
The Multipoint Calibration screen contains the following information:
●
Point 1:
First cal point used in the calibrated multipoint polynomial
curve.
●
Point 2:
Second cal point used in the calibrated multipoint polynomial
curve.
●
Point 3:
Third cal point used in the calibrated multipoint polynomial
curve.
●
First Coefficient:
Read only. Value of the first coefficient in the
polynomial curve.
●
Second Coefficient:
Read only. Value of the second coefficient in the
polynomial curve.
Multipoint Calibration
Summary of Contents for 410iQ
Page 1: ...410iQ Instruction Manual Carbon Dioxide Analyzer 117081 00 1Apr2020 ...
Page 2: ......
Page 18: ......
Page 35: ...Operation Instrument Display Thermo Scientific 410iQ Instruction Manual 3 9 ...
Page 160: ......
Page 171: ...Calibration Calibration Thermo Scientific 410iQ Instruction Manual 4 11 ...
Page 188: ......
Page 264: ......
Page 284: ......
Page 304: ......
Page 305: ......