Operation
Settings
3-102
410iQ Instruction Manual
Thermo Scientific
The Alarm Setpoints screen allows the user to view and set all settable
alarm minimum and maximum values. Alarm setpoints can also be set in
Settings>Health Check>Status and Alarms screens.
Note
User cannot set alarm limits outside of the acceptable range. The
minimum and maximum alarm limit can also be set by pressing on the
corresponding buttons located in the Settings>Health Check>Status and
Alarms screen. See “
Status and Alarms
” on page
3-45
.
▲
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Settings>Instrument Settings>Alarm Setpoints
The Alarm Setpoints screen contains the following information:
●
Across:
●
Item:
Lists the items that have settable alarm limits.
●
Value:
Displays the current value for each item.
●
Low Alarm:
User sets low alarm for item.
●
High Alarm:
User sets high alarm for item.
●
Units:
Units for each item (not settable).
●
Down:
●
CO2:
CO
2
concentration alarm.
●
Bkg Check Offset:
User can set the maximum allowable background
reading offset for calibration and calibration checks. This is set with
the high alarm only.
Alarm Setpoints
Summary of Contents for 410iQ
Page 1: ...410iQ Instruction Manual Carbon Dioxide Analyzer 117081 00 1Apr2020 ...
Page 2: ......
Page 18: ......
Page 35: ...Operation Instrument Display Thermo Scientific 410iQ Instruction Manual 3 9 ...
Page 160: ......
Page 171: ...Calibration Calibration Thermo Scientific 410iQ Instruction Manual 4 11 ...
Page 188: ......
Page 264: ......
Page 284: ......
Page 304: ......
Page 305: ......