Operation
Settings
Thermo Scientific
410iQ Instruction Manual
3-87
When O
2
correction is enabled, the following equation is applied to the
concentration value to generate a corrected value. The corrected value may
be output on the analog outputs, stored in logging memory, or output as
streaming data. The corrected values are not displayed on the Home Screen
on the front panel, only the non-corrected values.
2
9
.
20
2
9
.
20
*
2
@
O
O
C
O
C
meas
corr
corr
comp
Abbreviations: C is the measured concentration, O2
corr
is the oxygen
concentration that C is corrected to (this value is set by the user in the
Source O
2
Concentration Setpoint screen and defaults to 15%). O2
meas
is
the measured O
2
concentration, which is clamped at 20.8% to prevent
divide-by-zero errors. C
comp
is the corrected concentration that may be used
for analog outputs or datalogging.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Compensation>O
2
Correction (single range mode)
O
2
Correction
Summary of Contents for 410iQ
Page 1: ...410iQ Instruction Manual Carbon Dioxide Analyzer 117081 00 1Apr2020 ...
Page 2: ......
Page 18: ......
Page 35: ...Operation Instrument Display Thermo Scientific 410iQ Instruction Manual 3 9 ...
Page 160: ......
Page 171: ...Calibration Calibration Thermo Scientific 410iQ Instruction Manual 4 11 ...
Page 188: ......
Page 264: ......
Page 284: ......
Page 304: ......
Page 305: ......