![Thermo Scientific 410iQ Instruction Manual Download Page 202](http://html.mh-extra.com/html/thermo-scientific/410iq/410iq_instruction-manual_1105140202.webp)
Servicing
LCD Module Replacement
7-14
410iQ Instruction Manual
Thermo Scientific
Use the following procedure to replace the LCD module.
Equipment required:
Wrench, 1/4
1.
Turn instrument OFF and unplug the power cord.
2.
Gripping from the top corners of the front panel and pull outwards.
3.
Using 1/4 wrench, unfasten four #4-40 nuts (
Figure 7–15
).
Figure 7–15.
Replacing the LCD Module
4.
Remove cover.
5.
Unplug LCD cables from backside of board.
6.
Pull board off the standoffs.
LCD Module
Replacement