Servicing
DMC Optical Bench
7-42
410iQ Instruction Manual
Thermo Scientific
Use the following procedure to replace the filter wheel.
Equipment required:
Hex drive, 5/32 and 5/64
Phillips drive, #2
1.
Remove optical bench. Refer to “
Optical Bench Removal
” on page
7-
33
.
2.
Remove purge wheel housing. Refer to “
Purge Wheel Housing
Removal
” on page
7-40
.
3.
Using 5/32 hex drive, unfasten two #10-32 socket cap head screws.
4.
Using #2 Phillips drive, unfasten #8-32 pan head screw to gain access
to #8-32 set screw.
5.
Using 5/64 hex drive, loosen #8-32 set screw two or three turns only.
6.
Carefully slide chopper wheel off motor shaft.
7.
Replace filter wheel and align motor and filter wheel. Refer to “
Motor
and Filter Wheel Alignment
” on page
7-46
.
Filter Wheel
Replacement