Servicing
Optional DMC Oxygen Sensor
7-64
410iQ Instruction Manual
Thermo Scientific
Use the following to replace the oxygen sensor capillary.
Equipment required:
Ball-end hex drive, 3/32
1.
Remove the oxygen sensor DMC from the instrument. Refer to
“
Oxygen Sensor Removal
” on page
7-59
.
2.
Using a 3/32 ball-end hex drive (angle alignment of screws), unfasten
two #4-40 cap screws.
3.
Unlock tube clamp.
4.
Unplug tube after unlocking tube clamp.
Figure 7–73.
Replacing the Oxygen Sensor Capillary
Oxygen Sensor
Capillary
Replacement