![Thermo Scientific 410iQ Instruction Manual Download Page 226](http://html.mh-extra.com/html/thermo-scientific/410iq/410iq_instruction-manual_1105140226.webp)
Servicing
DMC Optical Bench
7-38
410iQ Instruction Manual
Thermo Scientific
Note
Replace the optical switch if needed. You don’t have to remove the
optical switch to replace the motor or chopper.
▲
5.
Using #1 Phillips drive, unfasten two #4-40 flat head screws to remove
the optical cover.
6.
Using #1 Phillips drive, unfasten one #2-56 pan head screw to remove
the optical switch.
Figure 7–44.
Optical Switch Replacement (pt 3) (Standard)