Servicing
Capillary Cleaning and/or Replacement
7-24
410iQ Instruction Manual
Thermo Scientific
Use the following procedure to clean and/or replace the capillary.
Equipment required:
Phillips drive, #2
Hex drive, 7/64
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Disconnect the plumbing.
3.
Using #2 Phillips drive, unfasten captive hardware.
Figure 7–27.
Remove Capillary, Disconnect and Unfasten
4.
Slide capillary plate upwards clearing the partition panel keyway.
Capillary
Cleaning and/or
Replacement