When do I need to open the spectrometer cover?
You need to remove the spectrometer cover to change sources or
detectors or install a spectrometer component such as a laser or
electronics module.
Warning
Do not operate the spectrometer with the main cover removed. The
cover protects you from exposure to laser light and live electrical
connections. If you must open the main cover for maintenance or
service reasons, always turn off the spectrometer power and unplug
the power cord first. If you have a CsI configuration, protect the
beamsplitter from the effects of water vapor by minimizing the time
that the cover is removed.
▲
In Spectrometer Help Topics find “Removing the main cover” in
“Servicing your spectrometer” help book
When should I align the spectrometer?
Align the spectrometer in the following situations:
●
The ambient temperature has changed by at least 5 Celsius
degrees (about 9 Fahrenheit degrees).
●
You are about to perform a spectrometer validation with Val-Q
(or other software). This ensures that the validation is run with
the best possible interferometer and electronics settings.
●
The spectrometer power was turned off for more than a few
minutes and has now been on for one hour.
Note
We recommend that you leave the power on continuously, but if you
do need to turn it off, allow the spectrometer to warm up for one
hour and then align it before collecting spectra.
▲
Nicolet 380 User’s Guide
91