Problem Possible
Cause
Solution
The spectrometer is overheated.
(You can check whether the
temperature of the electronics
board is within the acceptable
range by clicking the electronics
indicator on the Diagnostics tab of
the Experiment Setup dialog box.)
There is insufficient air flow around
the spectrometer.
Make sure the cooling vents on
the back of the spectrometer are
not blocked and nothing is
stacked on top of the
spectrometer. Maintain at least
30 cm (12 in) of clearance behind
the spectrometer and 64 cm
(25 in) of clearance above the
table.
Ambient temperature around the
spectrometer is too high.
Make sure the ambient
temperature is between 16° and
27°C (60° and 80°F).
An error message appears when
you try to scan
The light source is not working.
Check the source indicator on the
Diagnostic tab of the Experiment
Setup dialog box. If necessary,
replace the source.
An MCT detector is not cooled.
Check and cool the detector.
The spectrometer is out of
alignment.
Align the spectrometer using the
Align button on the Diagnostic
tab of the Experiment Setup
dialog box.
The interferogram peak is not
located at 1024 ± 16 data points on
the Bench tab of the Experiment
Setup dialog box.
Align the spectrometer. If the
interferogram is still not located
correctly, call Thermo Electron
for service.
The beam path is obstructed.
Remove the obstruction.
continued…
Nicolet 380 User’s Guide 113