Checking purge
If the spectrometer is equipped with an optional purge kit, make sure
the pressure regulator is set between 10 and 20 psi and the flowmeter
is set to 15 scfh.
Important
Excessive purge flow rates can introduce noise into the spectra. Do
not exceed 15 scfh.
▲
Important
CsI configurations must be purged at all times to protect the CsI
beamsplitter from being fogged by exposure to water vapor.
▲
Warning
Never use a flammable gas to purge the spectrometer. The purge gas
must be free of moisture, oil, carbon dioxide and other reactive or
infrared-absorbing materials. Use dried air or nitrogen to purge the
spectrometer. Other gases, even inert gases such as argon, can
damage the spectrometer and should never be used for this
purpose.
▲
We recommend that you leave the purge on at all times. This keeps
the spectrometer free of undesirable gases, protects the optics and
improves the system’s thermal stability.
In Spectrometer Help Topics view “Setting the purge gas controls”
in the “Using your spectrometer” book and “Checking and changing
the purge gas filter” in the “Maintaining your spectrometer” book for
information about how to install the purge equipment, set the
controls for the first time, and inspect and clean the purge filter.
Nicolet 380 User’s Guide 103