Problem Possible
Cause
Solution
The status indicators on the
spectrometer cover do not light
when the spectrometer is turned
on.
The power supply connection on the
back of the spectrometer is loose.
Turn off the spectrometer and
check the connections between
the spectrometer and the power
supply. Make sure the screws are
tightened finger-tight.
The spectrometer is not plugged in
Make sure the power cord is
securely connected to the power
supply and a working outlet or
power strip.
The power cord is not properly rated
for the electrical service.
Make sure the power cord is
appropriate for the AC power
source. In Spectrometer Help
Topics see “Miscellaneous parts”
in the “Ordering parts” book for
power cord descriptions and part
number information.
The ground prong on the power cord
has been removed or is defective.
The power cable or power supply is
defective.
Replace the power cord. In
Spectrometer Help Topics see
“Miscellaneous parts” in the
“Ordering parts” book for power
cord descriptions and part
number information.
The signal intensity is low.
(The Scan indicator may flash
intermittently and the
spectrometer alignment may fail.)
The source is glowing unevenly.
Check to see if the source
element is glowing evenly. Order
a replacement source if needed.
The beam path is obstructed.
Remove the obstruction.
continued…