Problem Possible
Cause
Solution
The signal intensity is low,
continued…
The sample compartment KBr
windows are fogged.
Call Thermo Electron to have the
windows replaced.
The beamsplitter is fogged.
Check the laser signals by
clicking the laser indicator on the
Diagnostic tab of the Experiment
Setup dialog box. If you suspect
the beamsplitter is fogged, call
Thermo Electron for service.
The spectrometer is out of
alignment.
Align the spectrometer using the
Align button on the Diagnostic
tab of the Experiment Setup
dialog box.
A sampling accessory is not
installed properly.
Make sure the accessory is
installed according to the
instructions that came with it.
A detector that was just installed is
not seated properly.
Make sure the hole and slot in the
detector plate fit over the pins on
the spectrometer baseplate. See
“Replacing the detector” in the
“Servicing your spectrometer”
book in Spectrometer Help
Topics.
The system scans normally but
gives very low signal intensity.
The spectrometer is out of
alignment.
Align the spectrometer using the
Align button on the Diagnostic
tab of the Experiment Setup
dialog box.
continued…
Nicolet 380 User’s Guide 111