Problem Possible
Cause
Solution
The system scans normally but
gives very low signal intensity,
continued…
The moving mirror velocity is set
too high.
Set Velocity on the Bench tab of
the Experiment Setup dialog box
to a lower velocity.
There is no interferogram.
If the Source status indicator is
off, click the source indicator on
the Diagnostic tab of the
Experiment Setup dialog box to
check the source current and
voltage. If the Source indicator is
on, check the interferogram on
the Bench tab of the Experiment
Setup dialog box; if the
interferogram is not present,
check the beam path and
detector.
The beam path is obstructed
Remove the obstruction.
The automated aperture is set
incorrectly
Check the Aperture parameter on
the Bench tab of the Experiment
Setup dialog box. For an MCT-A
detector, set Aperture to 10. For a
DTGS detector, set Aperture to
100.
A sampling accessory is not
installed properly or is not aligned.
Make sure the accessory is
installed correctly and, if
necessary, align it. See the
instructions that came with the
accessory.
continued…
Thermo Electron Corporation
112