Operating Instructions
Teledyne API - Model T200H/T200M Operation Manual
82
Table 4-9: T200H/M Default DAS Configuration
CHANNELS with PROPERTIES
PARAMETERS
NAME
MODE
EVENT
PRECISION
NUM
SAMPLES
Name: CONC
Event: ATIMER
Sample Period: 000:00:01
Report Period: 000:01:00
Number of Records: 800
RS-232 report: OFF
Channel enabled: ON
DAS HOLDOFF: ON
NOXCNC1
AVG
- -
4
ON
NOCNC1
AVG
- -
4
OFF
N2CNC1
AVG
- -
4
OFF
STABIL
AVG
- -
4
OM
Name: CALDAT
Event: SLPCHG
Number of Records: 200
RS-232 report: OFF
Channel enabled: ON
DAS HOLDOFF: OFF
NXZSC1
- -
SLPCHG
4
OFF
NOXSLP1
- -
SLPCHG
4
OFF
NOXOFFS1
- -
SLPCHG
4
OFF
NOZSC1
- -
SLPCHG
4
OFF
NOSLP1
- -
SLPCHG
4
OFF
NOOFFS1
- -
SLPCHG
4
OFF
N2ZSC1
- -
SLPCHG
4
OFF
CNVEF1
- -
SLPCHG
4
OFF
STABIL
- -
SLPCHG
4
OFF
Name: CALCHECK
Event: EXITMP
Number of Records: 200
RS-232 report: OFF
Channel enabled: ON
DAS HOLDOFF: OFF
NXZSC1
- -
EXITMP
4
OFF
NOZSC1
- -
EXITMP
4
OFF
N2ZSC1
- -
EXITMP
4
OFF
STABIL
- -
EXITMP
4
OFF
Name: CALCHECK
Event: EXITMP
Number of Records: 200
RS-232 report: OFF
Channel enabled: ON
DAS HOLDOFF: OFF
SMPFLW
AVG
- -
2
OFF
O3FLOW
AVG
- -
2
OFF
RCPRESS
AVG
- -
2
OFF
SMPPRES
AVG
- -
2
OFF
RCTEMP
AVG
- -
2
OFF
PMTTMP
AVG
- -
2
OFF
CNVTMP
AVG
- -
2
OFF
BOXTMP
AVG
- -
2
OFF
HVPS
AVG
- -
2
OFF
AZERO
AVG
- -
2
OFF
Name: HIRES
Event: ATIMER
Sample Period: 000:00:01
Report Period: 000:00:01
Number of Records: 1500
RS-232 report: OFF
Channel enabled: OFF
DAS HOLDOFF: OFF
NOXCNC1
AVG
- -
4
OFF
NOCNC1
AVG
- -
4
OFF
N2CNC1
AVG
- -
4
OFF
STABIL
AVG
- -
4
OFF
SMPFLW
AVG
- -
2
OFF
O3FLOW
AVG
- -
2
OFF
RCPRESS
AVG
- -
2
OFF
SMPPRES
AVG
- -
2
OFF
RCTEMP
AVG
- -
2
OFF
PMTTMP
AVG
- -
2
OFF
CNVTMP
AVG
- -
2
OFF
BOXTMP
AVG
- -
2
OFF
HVPS
AVG
- -
1
OFF
AZERO
AVG
- -
2
OFF
REFGND
AVG
1
OFF
REF4096
AVG
1
OFF
07270D DCN7141