Teledyne API - Model T200H/T200M Operation Manual
Calibration Procedures
185
STEP TWO:
Set the expected NO
2
span gas concentration:
SAMPLE
GAS TO CAL:NOX
NOX
O2
ENTR
EXIT
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST >
CAL
SETUP
M-P CAL A1:NXCNC1 =100PPM NOX=X.XXX
<TST TST> ZERO SPAN
CONC
EXIT
SAMPLE
RANGE TO CAL:LOW
LOW
HIGH
ENTR
EXIT
M-P CAL CONCENTRATION MENU
NOX
NO
CONV
EXIT
M-P CAL NO2 CE CONC:80.0 Conc
0
0
8
0
.0
ENTR EXIT
The NO
X
& NO span concentration
values automatically default to
80.0 Conc
.
If this is not the the concentration of
the span gas being used, toggle
these buttons to set the correct
concentration of the NO
X
and NO
calibration gases.
EXIT
ignores the new
setting and returns to
the previous display.
ENTR
accepts the new
setting and returns to
the
CONVERTER
EFFICIENCY
MENU.
If using NO span gas
in addition to NO
X
repeat last step.
M-P CAL CONVERTER EFFICIENCY MENU
NO2
CAL
SET
EXIT
07270D DCN7141