T200H/M and 200EH/EM Menu Trees (05147J DCN7141)
APPENDIX A-3: Warnings and Test Measurements
A-11
Table A-3: Test Measurements
Test Measurement Name
Message Text
Description
Test measurements
NONOXCONC
NO=396.5 NOX=396.5
3
Simultaneously displays NO and NOX
concentrations.
RANGE
not 6
RANGE=500.0 PPB
3
D/A range in single or auto-range modes.
RANGE1
not 6
RANGE1=500.0 PPB
3
D/A #1 range in independent range mode.
RANGE2
not 6
RANGE2=500.0 PPB
3
D/A #2 range in independent range mode.
RANGE3
not 6
RANGE3=500.0 PPB
3
D/A #3 range in independent range mode.
STABILITY
NOX STB=0.0 PPB
3
O2 STB=0.0 PCT
5
CO2 STB=0.0 PCT
8
Concentration stability (standard deviation
based on setting of
STABIL_FREQ
and
STABIL_SAMPLES
). Select gas with
STABIL_GAS
variable.
RESPONSE
2
RSP=8.81(1.30) SEC
Instrument response. Length of each
signal processing loop. Time in
parenthesis is standard deviation.
SAMPFLOW
SAMP FLW=460 CC/M
Sample flow rate.
OZONEFLOW
OZONE FL=87 CC/M
Ozone flow rate.
PMT
PMT=800.0 MV
Raw PMT reading.
NORMPMT
NORM PMT=793.0 MV
PMT reading normalized for temperature,
pressure, auto-zero offset, but not range.
AUTOZERO
AZERO=1.3 MV
Auto-zero offset.
HVPS
HVPS=650 V
High voltage power supply output.
RCELLTEMP
RCELL TEMP=50.8 C
Reaction cell temperature.
BOXTEMP
BOX TEMP=28.2 C
Internal chassis temperature.
REMBOXTEMP
10
REM BOX TMP=30.1 C
Remote chassis temperature.
PMTTEMP
PMT TEMP=7.0 C
PMT temperature.
MANIFOLDTEMP
4
MF TEMP=50.8 C
Bypass or dilution manifold temperature.
CO2CELLTEMP
8
CO2 CELL TEMP=50.8 C
CO
2
sensor cell temperature.
O2CELLTEMP
5
O2 CELL TEMP=50.8 C
O
2
sensor cell temperature.
IZSTEMP
IZS TEMP=50.8 C
IZS temperature.
CONVTEMP
MOLY TEMP=315.0 C
Converter temperature. Converter type is
MOLY
,
CONV
, or
O3KL
.
SAMPRESTTEMP
10
SMP RST TMP=49.8 C
Sample restrictor temperature.
RCELLPRESS
RCEL=7.0 IN-HG-A
Reaction cell pressure.
SAMPPRESS
SAMP=29.9 IN-HG-A
Sample pressure.
NOXSLOPE
NOX SLOPE=1.000
NOX slope for current range, computed
during zero/span calibration.
NOXOFFSET
NOX OFFS=0.0 MV
NOX offset for current range, computed
during zero/span calibration.
NOSLOPE
NO SLOPE=1.000
NO slope for current range, computed
during zero/span calibration.
NOOFFSET
NO OFFS=0.0 MV
NO offset for current range, computed
during zero/span calibration.
07270D DCN7141