Appendix C
Warranty/Repair Questionnaire
T200H/M, M200EH/EM
(05149C DCN6055)
TELEDYNE API CUSTOMER SERVICE
PHONE: (858) 657-9800 TOLL FREE: (800) 324-5190 FAX: (858) 657-9816
C-1
CUSTOMER: _____________________________________
PHONE: ________________________________
CONTACT NAME: _________________________________
FAX NO. ________________________________
SITE ADDRESS: _____________________________________________________________________________
MODEL TYPE: _______________ SERIAL NO.: _________________ FIRMWARE REVISION: ___________
1. Are there any failure messages? _______________________________________________________________
___________________________________________________________________________________________
___________________________________________________________________________________________
_________________________________________________________________ (Continue on back if necessary)
PLEASE COMPLETE THE FOLLOWING TABLE:
TEST FUNCTION
RECORDED VALUE
UNITS
ACCEPTABLE VALUE
NOx STAB
PPB/PPM
≤
1 PPB WITH ZERO AIR
SAMPLE FLOW
cm
3
/min
T200M/200EM: 250 ± 25
T200H/200EH: 300 ± 30
OZONE FLOW
cm
3
/min
250 ± 25
PMT SIGNAL WITH ZERO AIR
MV
-20 to 150
PMT SIGNAL AT SPAN GAS CONC
MV
PPB
0-5000MV
0-5,000 PPM
1
, 200 PPM
2
NORM PMT SIGNAL AT SPAN
GAS CONC
MV
PPB
0-5000MV
0-5,000 PPM
1
, 200 PPM
2
AZERO
MV
-20 to 150
HVPS
V
400 to 900
RCELL TEMP
ºC
50 ± 1
BOX TEMP
ºC
AMBIENT ± 5ºC
PMT TEMP
ºC
7 ± 2ºC
O
2
CELL TEMP
1
ºC
30ºC to 70ºC
IZS TEMP
1
ºC
50 ± 1ºC
MOLY TEMP
ºC
315 ± 5ºC
RCEL (pressure)
IN-HG-A
<10
SAMP (pressure)
IN-HG-A
AMBIENT ± 1
NOx SLOPE
1.0 ± 0.3
NOx OFFSET
mV
50 to 150
NO SLOPE
1.0 ± 0.3
NO OFFSET
mV
50 to 150
O2 SLOPE
1
0.5 to 2.0
O2 OFFSET
1
%
-10 to + 10
PMT SIGNAL DURING ETEST
MV
2000 ± 1000
PMT SIGNAL DURING OTEST
MV
2000 ± 1000
REF_4096_MV
2
MV
4096mv ±2mv and Must be Stable
REF_GND
2
MV
0± 0.5 and Must be Stable
1
If option is installed
2
Located in Signal I/O list under
DIAG
menu