Teledyne API - Model T200H/T200M Operation Manual
Operating Instructions
147
4.13.6.1. Automatic Analog Output Calibration
To calibrate the outputs as a group with the
AOUTS CALIBRATION
command, press:
From the
AIO CONFIGURATION SUBMENU
(in the SETUP DIAG menu)
Analyzer
automatically
calibrates all
channels for which
AUTO-CAL
is turned
ON
If any of the channels
have not been calibrated
ot if at least one channel
has AUTO-CAL turned
OFF, this message will
read
NO
.
DIAG AIO
AUTO CALIBRATING DATA_OUT_1
DIAG AIO
AUTO CALIBRATING DATA_OUT_2
DIAG AIO
NOT AUTO CAL. DATA_OUT_3
DIAG AIO
AUTO CALIBRATING DATA_OUT_4
This message
appears when
AUTO-CAL
is
Turned
OFF
for
a channel
DIAG AIO AOUTS CALIBRATED: YES
SET> CAL
EXIT
DIAG AIO AOUTS CALIBRATED: NO
SET>
CAL
EXIT
DIAG
ANALOG I/O CONFIGURATION
PREV NEXT
ENTR
EXIT
Note
Manual calibration should be used for the 0.1V range or in cases where the
outputs must be closely matched to the characteristics of the recording device.