T200H/M and 200EH/EM Menu Trees (05147J DCN7141)
APPENDIX A-4: M Signal I/O Definitions
A-15
Signal Name
Bit or Channel
Number
Description
Control outputs, U21, J1008, pins 9
–
12 = bits 0
–
3, default I/O address 325 hex
0–3
Spare
Alarm outputs, U21, J1009, pins 1
–
12 = bits 4
–
7, default I/O address 325 hex
ST_SYSTEM_OK2
12
4
1 = system OK
0 = any alarm condition or in diagnostics mode
MB_RELAY_36
18
Controlled by MODBUS coil register
OUT_CAL_MODE
13
1 = calibration mode
0 = measure mode
ST_CONC_ALARM_1
17
5
1 = conc. limit 1 exceeded
0 = conc. OK
MB_RELAY_37
18
Controlled by MODBUS coil register
OUT_SPAN_CAL
13
1 = span calibration
0 = zero calibration
ST_CONC_ALARM_2
17
6
1 = conc. limit 2 exceeded
0 = conc. OK
MB_RELAY_38
18
Controlled by MODBUS coil register
OUT_PROBE_1
13
0 = select probe #1
1 = not selected
ST_HIGH_RANGE2
19
7
1 = high auto-range in use (mirrors
ST_HIGH_RANGE
status output)
0 = low auto-range
MB_RELAY_39
18
Controlled by MODBUS coil register
OUT_PROBE_2
13
0 = select probe #2
1 = not selected
A status outputs, U24, J1017, pins 1
–
8 = bits 0
–
7, default I/O address 323 hex
ST_SYSTEM_OK
0
0 = system OK
1 = any alarm condition
ST_CONC_VALID
1
0 = conc. valid
1 = conc. filters contain no data
ST_HIGH_RANGE
2
0 = high auto-range in use
1 = low auto-range
ST_ZERO_CAL
3
0 = in zero calibration
1 = not in zero
ST_SPAN_CAL
4
0 = in span calibration
1 = not in span
ST_DIAG_MODE
5
0 = in diagnostic mode
1 = not in diagnostic mode
ST_LOW_SPAN_CAL
20
6
0 = in low span calibration
1 = not in low span
ST_O2_CAL
11
7
0 = in O
2
calibration mode
1 = in measure or other calibration mode
07270D DCN7141