Teledyne API - Model T200H/T200M Operation Manual
Getting Started
67
STEP 4 – O
2
ZERO/SPAN CALIBRATION :
To perform the zero/span calibration procedure:
Press
ENTR
to changes
the
OFFSET
&
SLOPE
values for the O
2
measurement.
Press
EXIT
to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE
GAS TO CAL:NOX
NOX
O2
ENTR
EXIT
SAMPLE
O2 STB=X.XX %
NOX=XXX.X
< TST TST > CAL SETUP
M-P CAL
O2 STB=X.XX %
NOX=XXX.X
<TST TST>
ZERO
CONC
EXIT
SAMPLE
RANGE TO CAL:LOW
LOW
HIGH
ENTR
EXIT
EXIT
at this point
returns to the
SAMPLE menu.
Press
ENTR
to changes
the
OFFSET
&
SLOPE
values for the O
2
measurement.
Press
EXIT
to leave the
calibration unchanged and
return to the previous
menu.
Allow zero gas to enter the sample port
at the rear of the analyzer.
Wait until
O2 STB
falls below
0.1%.
This may take several
minutes.
SAMPLE
O2 STB=X.XX %
NOX=XXX.X
< TST TST >
CAL
SETUP
M-P CAL
O2 STB=X.XX %
NOX=XXX.X
<TST TST>
ENTR
CONC
EXIT
SAMPLE
GAS TO CAL:NOX
NOX
O2
ENTR EXIT
M-P CAL
O2 STB=X.XX %
NOX=XXX.X
<TST TST> ZERO
SPAN
CONC
EXIT
SAMPLE
RANGE TO CAL:LOW
LOW
HIGH
ENTR
EXIT
Allow span gas to enter the sample port
at the rear of the analyzer.
Wait until
O2 STB
falls below
0.1%
This may take several
minutes.
SAMPLE
O2 STB=X.XX %
NOX=XXX.X
< TST TST >
CAL
SETUP
M-P CAL
O2 STB=X.XX %
NOX=XXX.X
<TST TST>
ENTR
CONC
EXIT
M-P CAL
O2 STB=X.XX %
NOX=XXX.X
<TST TST> ENTR
CONC
EXIT
The
SPAN
button now
appears during the transition
from zero to span.
You may see both keys.
If either the
ZERO
or
SPAN
buttons fail to appear see
Section 10 for
troubleshooting tips.
Analyzer continues to
cycle through NOx,
NO, NO
2
and O
2
measurements
throughout this
procedure.
SAMPLE A1:CONC1=50PPM
NOX=XXXX
< TST
TST >
CAL SETUP
Set the Display to show
the
O2 STB
test function.
This function calculates
the stability of the O
2
measurement
Toggle
TST>
button until ...
SAMPLE
GAS TO CAL:O2
NOX
O2
ENTR
EXIT
The Model T200H/M analyzer is now ready for operation.