Teledyne API - Model T200H/T200M Operation Manual
Operating Instructions
133
4. Choose an DAS parameter to be output on the channel.
5. Set the reporting range scale for the data type chosen.
6. Set the update rate for the channel.
7. Calibrating the output channel. This can be done automatically or manually for
each channel (see Section 4.13.6).
To access the analog I/O configuration sub menu, press:
AIO Configuration Submenu
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG
DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
ALRM
EXIT
SETUP X.X
ENTER PASSWORD:818
8
1
8
ENTR
EXIT
DIAG SIGNAL I/O
NEXT
ENTR
EXIT
Continue pressing
NEXT
until ...
DIAG
ANALOG I/O CONFIGURATION
PREV NEXT
ENTR
EXIT
DIAG AIO A OUTS CALIBRATED: NO
<SET
SET>
CAL
EXIT
DIAG AIO DATA_OUT_1: 5V, NXCNC1, NOCAL
<SET
SET>
EDIT
EXIT
DIAG AIO AIN CALIBRATED: NO
<SET
SET>
CAL
EXIT
DIAG AIO DATA_OUT_2: 5V, NXCNC1, NOCAL
<SET
SET>
EDIT
EXIT
DIAG AIO DATA_OUT_3: 5V, NXCNC1, NOCAL
<SET
SET>
EDIT
EXIT
DIAG AIO DATA_OUT_4: 5V, NXCNC1, NOCAL
<SET
SET>
EDIT
EXIT
07270D DCN7141