Teledyne API - Model T200H/T200M Operation Manual
Operating Instructions
113
Internet Configuration Button Functions
BUTTON
FUNCTION
[0]
Location of cursor. Press to cycle through the range of
numerals and available characters (“
0 – 9”
&
“ . ”
)
<CH CH>
Moves the cursor one character left or right.
DEL
Deletes a character at the cursor location.
ENTR
Accepts the new setting and returns to the previous
menu.
EXIT
Ignores the new setting and returns to the previous
menu.
Some buttons appear only when relevant.
SETUP X.X
DHCP: OFF
SET>
EDIT
EXIT
SETUP X.X
INST IP: 000.000.000.000
<SET SET> EDIT
EXIT
SETUP X.X
GATEWAY IP: 000.000.000.000
<SET SET> EDIT
EXIT
SETUP X.X
INST IP: [0] 00.000.000
<CH CH> DEL [0]
ENTR EXIT
SETUP X.X
SUBNET MASK:255.255.255.0
<SET SET> EDIT
EXIT
SETUP X.X
SUBNET MASK:[2]55.255.255.0
<CH CH> DEL [?]
ENTR EXIT
SETUP X.X
TCP PORT 3000
<SET
EDIT
EXIT
The PORT number must remain at
3000
.
Do not change this setting unless instructed to by
Teledyne Instruments Customer Service personnel.
SETUP X.X
GATEWAY IP: [0] 00.000.000
<CH CH> DEL [?]
ENTR EXIT
Cursor
location is
indicated by
brackets
SETUP X.X
INITIALIZING INET 0%
…
INITIALIZING INET 100%
SETUP X.X
INITIALIZATI0N SUCCEEDED
SETUP X.X
INITIALIZATION FAILED
SETUP X.X
COMMUNICATIONS MENU
ID
INET
COM1
EXIT
Pressing
EXIT
from
any of the above
display menus
causes the Ethernet
option to reinitialize
its internal interface
firmware
Contact your IT
Network Administrator
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
COMMUNICATIONS MENU
ID
INET
COM1
EXIT
DHCP: ON is
default setting.
Skip this step
if it has been
set to OFF.
SETUP X.X
DHCP: ON
SET>
EDIT
EXIT
ENTR
accepts
the new
assigned
numbers;
EXIT
ignores
Figure 4-7:
COM – LAN / Internet Manual Configuration
07270D DCN7141