APPENDIX A-4: Signal I/O Definitions
Teledyne API - T300/T300M and M300E/EM 04906J DCN7122
A-16
Signal Name
Bit or Channel
Number
Description
Front panel I
2
C keyboard, default I
2
C address 4E hex
MAINT_MODE
5 (input)
0 = maintenance mode
1 = normal mode
LANG2_SELECT
6 (input)
0 = select second language
1 = select first language (English)
SAMPLE_LED
8 (output)
0 = sample LED on
1 = off
CAL_LED
9 (output)
0 = cal. LED on
1 = off
FAULT_LED
10 (output)
0 = fault LED on
1 = off
AUDIBLE_BEEPER
14 (output)
0 = beeper on (for diagnostic testing only)
1 = off
Relay board digital output (PCF8574), default I
2
C address 44 hex
RELAY_WATCHDOG
0
Alternate between 0 and 1 at least every 5 seconds to keep
relay board active
WHEEL_HTR
1
0 = wheel heater on
1 = off
BENCH_HTR
2
0 = optical bench heater on
1 = off
O2_CELL_HEATER
5
3
0 = O
2
sensor cell heater on
1 = off
BOX2_HEATER
3
,
OVEN_HEATER
15
3
0 = internal box temperature #2/oven heater on
1 = off
CAL_VALVE
4
0 = let cal. gas in
1 = let sample gas in
SPAN_VALVE
5
0 = let span gas in
1 = let zero gas in
ZERO_SCRUB_VALVE
2,3
6
0 = open zero scrubber valve
1 = close
SHUTOFF_VALVE
6
7
3,15
0 = energize shutoff valve
1 = de-energize
IR_SOURCE_ON
7
n/a
3,15
0 = IR source on
1 = off
06864D DCN7562