To adjust the signal levels of an analog output channel manually, select the
ANALOG
I/O CONFIGURATION
submenu (see Figure 5-3) then press:
From the
AIO CONFIGURATION SUBMENU
DIAG
ANALOG I/O CONFIGURATION
PREV NEXT
ENTR
EXIT
Continue pressing
SET>
until you reach the
output to be configured
DIAG AIO CONC_OUT_2: 5V, CONC2, NOCAL
<SET
SET>
EDIT
EXIT
Continue pressing
SET>
until ...
DIAG AIO CONC_OUT_2: CALIBRATED:NO
<SET
SET>
CAL
EXIT
DIAG AIO CONC_OUT_2: RANGE: 5V
SET>
EDIT
EXIT
DIAG AIO CONC_OUT_2: VOLT-Z: 0 mV
U100 UP10 UP DOWN DN10 D100 ENTR
EXIT
DIAG AIO CONC_OUT_2: VOLT-S: 4500 mV
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
DIAG AIO CONC_OUT_2: CALIBRATED: YES
<SET
SET> CAL
EXIT
These menus
only appear if
AUTO-CAL
is
turned
OFF.
These buttons increase /
decrease the analog output
signal level (not the value on the
display)
by 100, 10 or 1 counts.
Continue adjustments until the
voltage measured at the output
of the analyzer and/or the input
of the recording device matches
the value in the upper right hand
corner of the display
(within the tolerances
listed in Table 8.7.)
DIAG AIO AOUTS CALIBRATED: NO
SET>
CAL
EXIT
DISPLAYED AS
= CHANNEL
CONC_OUT_1
=
A1
CONC_OUT_2
=
A2
CONC_OUT_3
=
A3
TEST OUTPUT
=
A4
CONC_OUT_3
is only active on
analyzers with the
optional CO
2
or O
2
sensor installed.