Teledyne API – Model T300/T300M CO Analyzer
Getting Started
73
While these are the default setting for the T300/T300M Analyzer, it is recommended
that you verify them before proceeding with the calibration procedure, by pressing:
SETUP X.X
RANGE MODE:SINGL
SNGL DUAL AUTO
ENTR
EXIT
SAMPLE RANGE=50.0 PPM CO= XX.XX
<TST
TST>
CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
DIL
EXIT
SETUP X.X
RANGE: 50.0 Conc
0
0
0
5
0
.0
ENTR
EXIT
SETUP X.X
CONC UNITS:PPM
PPB
PPM
UGM MGM
ENTR
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
DIL
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
DIL
EXIT
Verify that the
MODE
is set for
SNGL.
If it is not, press
SINGL
à
ENTR.
Press
EXIT
as
needed to
return to
SAMPLE
m
ode.
Verify that the
RANGE
is
set for
50.0
If it is not, toggle each
numeric button until the
proper range is set, then
press
ENTR
.
Verify that the
UNIT
is set for
PPM
If it is not, press
PPM
à
ENTR.