APPENDIX A-3: Warnings and Test Functions
Teledyne API - T300/T300M and M300E/EM 04906J DCN7122
A-12
TEST FUNCTION NAME
MESSAGE TEXT
DESCRIPTION
COOFFSET2
OFFSET2=0.000
CO OFFSET2=0.000
7
CO offset for range #2 in dual range
mode, computed during zero/span
calibration.
CO2SLOPE
7
CO2 SLOPE=1.000
CO
2
slope, computed during zero/span
calibration.
CO2OFFSET
7
CO2 OFFSET=0.000
CO
2
offset, computed during zero/span
calibration.
O2SLOPE
10
O2 SLOPE=0.980
O
2
slope, computed during zero/span
calibration.
O2OFFSET
10
O2 OFFSET=1.79 %
O
2
offset, computed during zero/span
calibration.
CO
CO=17.7 PPM
3
CO concentration for current range.
CO2
7
CO2=15.0 %
CO
2
concentration.
O2
10
O2=0.00 %
O
2
concentration.
TESTCHAN
TEST=1751.4 MV
Value output to
TEST_OUTPUT
analog
output, selected with
TEST_CHAN_ID
variable.
CLOCKTIME
TIME=09:52:20
Current instrument time of day clock.
RANGE(s)
User Configurable
CO2RANGE
CO2 RANGE=20 %
1
CO
2
range.
O2RANGE
O2 RANGE=100 %
2
O
2
range.
STABILITY
STABIL=0.0 PPM
CO STB=0.0 PPM
1, 2
CO2 STB=0.0 %
1
O2 STB=0.0 %
2
Concentration stability (standard deviation
based on setting of
STABIL_FREQ
and
STABIL_SAMPLES
).
COMEAS
CO MEAS=4125.0 MV
Detector measure reading.
COREF
CO REF=3750.0 MV
Detector reference reading.
MRRATIO
MR RATIO=1.100
Measure/reference ratio.
SAMPPRESS
PRES=29.9 IN-HG-A
Sample pressure.
SAMPFLOW
SAMP FL=751 CC/M
Sample flow rate.
SAMPTEMP
SAMPLE TEMP=26.8 C
Sample temperature.
BENCHTEMP
BENCH TEMP=48.1 C
Bench temperature.
WHEELTEMP
WHEEL TEMP=68.1 C
Wheel temperature.
O2CELLTEMP
2
O2 CELL TEMP=50.2 C
O
2
sensor cell temperature.
BOXTEMP
BOX TEMP=26.8 C
Internal chassis temperature.
PHOTOTEMP
PHT DRIVE=2500.0 MV
Photometer temperature.
COSLOPE
SLOPE=1.000
CO SLOPE=1.000
1
CO slope for current range, computed
during zero/span calibration.
COSLOPE1
SLOPE1=1.000
CO SLOPE1=1.000
1
CO slope for range #1 in dual range
mode, computed during zero/span
calibration.
COSLOPE2
SLOPE2=1.000
CO SLOPE2=1.000
1
CO slope for range #2 in dual range
mode, computed during zero/span
calibration.
COOFFSET
OFFSET=0.000
CO OFFSET=0.000
1
CO offset for current range, computed
during zero/span calibration.
COOFFSET1
OFFSET1=0.000
CO OFFSET1=0.000
1
CO offset for range #1 in dual range
mode, computed during zero/span
calibration.
COOFFSET2
OFFSET2=0.000
CO OFFSET2=0.000
1
CO offset for range #2 in dual range
mode, computed during zero/span
calibration.
CO2SLOPE
1
CO2 SLOPE=1.000
CO
2
slope, computed during zero/span
calibration.
CO2OFFSET
1
CO2 OFFSET=0.000
CO
2
offset, computed during zero/span
calibration.
06864D DCN7562