Teledyne API – Model T300/T300M CO Analyzer
Setup Menu
105
To access and navigate the
VARS
menu, use the following button sequence.
SETUP X.X 4) CLOCK_ADJUST=0 Sec/Day
PREV
NEXT
JUMP
EDIT ENTR EXIT
SETUP X.X CLOCK_ADJUST=0 Sec/Day
+
0
0
ENTR EXIT
Enter sign and number of
seconds per day the clock
gains (-) or loses(+).
SETUP X.X
DYN_SPAN=OFF
OFF
ENTR EXIT
SETUP X.X
DYN_ZERO=OFF
OFF
ENTR EXIT
Toggle this button to turn
the Dynamic Zero
calibration feature
ON/
OFF.
Concentration display
continuously cycles
through all gasses.
SAMPLE
RANGE=500.0 PPM CO= XXXX
<TST
TST> CAL
SETUP
SETUP X.X
0) DAS_HOLD_OFF=15.0 Minutes
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
EXIT
SETUP X.X
ENTER PASSWORD:818
8
1
8
ENTR
EXIT
Toggle these
buttons to enter
the correct
PASSWORD.
In all cases:
EXIT
discards the new
setting.
ENTR
accepts the
new setting.
SETUP X.X
2) DYN_ZERO=OFF
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
3) DYN_SPAN=OFF
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
DAS_HOLD_OFF=15.0 Minutes
1
5
.0
ENTR EXIT
Toggle these buttons to set
the iDAS HOLDOFF time
period in minutes
(MAX = 20 minutes).
Toggle this button to turn
the Dynamic Span
calibration feature
ON/
OFF.
SETUP X.X
1) CONC_PRECISION=AUTO
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
CONC_PRECISION=AUTO
AUTO
1
2
3
4
ENTR EXIT
SETUP X.X
5) STABIL_GAS=CO
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
STABIL_GAS=O2
CO
CO2
O2
ENTR EXIT
Use these buttons to select
which gas will be reported
by the sTABIL test
function.
(O
2
is only available if the
optional O
2
sensor is
installed)
Use these buttons to select
the precision of the o33
concentration display.
Press
NEXT
for additional
VARS; press
NEXT
or
PREV
to move back and
forth throughout the list of
VARS.
06864D DCN7562